ClearanceBrief
Recall Z-1668-2020 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Biomet, Inc.

BIOMET StageOne Knee Tibial Cement Spacer Mold, 75 MM, Silicone, Sterile, Item 433175.

DateApr 3, 2020 (initiated Feb 26, 2020)
StatusTerminated
ReasonPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Root causeProcess control
ActionOn 2/26/2020, Zimmer Biomet sent an "Urgent Medical Device Recall" letter via FEDEX to direct sales distributors, hospitals, and surgeons. In addition, an email was sent to all distributors alerting them of the recall. Customers were instructed to quarantine the product and a Zimmer Biomet representative would remove the affected product from the facility. They asked customers to complete and return the "Certificate of Acknowledgement" form indicating whether they had any of the affected lots.
Quantity230,001
DistributionWorldwide distribution. US distribution to the following states: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Miss
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050210DISPOSABLE CEMENT SPACER MOLDS FOR TEMPORARY KNEE PROSTHESES
Biomet, Inc., Jul 19, 2005
K161273StageOne Disposable Cement Spacer Molds for Temporary Knee Prosthesis
Biomet, Inc., Jan 25, 2017