Recall Z-1954-2020 ยท KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type

Cardiac Assist, Inc

TandemHeart pump is assembled into kits: TandemLung Kit - DL31, Product: 5730-3118 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.

DateMar 6, 2020
StatusTerminated
ReasonFailure to prime due to an assembly error
Root causeUnder Investigation by firm
ActionLivaNova dba Tandem Life notified consignees via 1st class mail (certified return receipt) on March 06th, 2020, states reason for recall, health risk, and action to take: Check your inventory for the presence of potentially affected TandemHeart pump contained within your ordered kit as per Table 1 provided in Attachment 1; - Ensure any affected kit containing the potentially affected TandemHeart pump still in inventory gets quarantined before they can be returned. 2. Please fill and return Response Form by email to LivaNova.FSCA@livanova.comto confirm appropriate corrective actions have been
Quantity2
DistributionUS Nationwide distributions.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110493TANDEMHEART PUMP
Cardiacassist, Inc., Sep 20, 2011