Recall Z-2293-2020 ยท PGY Display, Diagnostic Radiology

Eizo Corporation

RadiForce GX560 monochrome LCD monitor, models 0FTD2347 and 0FTD2379 - Product Usage: is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

DateJul 30, 2019
StatusOpen, Classified
ReasonUnder certain conditions, a marble pattern infrequently appears on the monitor.
Root causeDevice Design
ActionEIZO Japan announced the recall on their website on 7/30/2019. EIZO U.S. began contacting their consignees via email and phone (one as early as 8/7/2019, some on 10/18/2019, but the majority on 11/26/2019) and began updating consignee monitors with the firmware. The notification requested the consignee contact EIZO U.S. for assistance with the firmware update.
Quantity96 monitors dist. to the U.S.
DistributionWorldwide distribution - US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Kansas, Louisiana, Maine, Massachusetts, Michigan, Missouri, Montana, Nevada, New Hampshire, New Jersey, New Me