Recall Z-1834-2020 · KLI Enzyme Immunoassay, Non-Radiolabeled, Total Thyroxine

Ortho Clinical Diagnostics

VITROS Immunodiagnostic Products TT4 Reagent Pack, Product Code 8744468

DateMar 10, 2020
StatusTerminated
ReasonThe specified reagent packs exhibit an increase in the frequency of unsuccessful calibrations.
Root causeUnder Investigation by firm
ActionIMPORTANT PRODUCT CORRECTION NOTIFICATION letters dated 3/10/20 were sent to customers. REQUIRED ACTIONS " If you have VITROS TT4 Reagent Packs Lot 2340 in your inventory, Ortho will replace or credit your account as indicated on your Confirmation of Receipt form. NOTE: As a reminder, if you have a successful calibration, and quality control results are acceptable, it is acceptable to continue using the affected lot. " Post this notification by each system that processes VITROS TT4 Reagent Packs, Lot 2340. " Complete the Confirmation of Receipt form now to register receipt of the notification
Quantity3181
DistributionUS Nationwide. Australia, Bermuda, Belgium, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160495VITROS® Automation Solutions
Ortho-Clinical Diagnostics, Inc., Oct 18, 2016