Recall Z-1796-2020 ยท DXC Clamp, Vascular

Merit Medical Systems, Inc.

PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT

DateMar 11, 2020
StatusTerminated
ReasonDuring the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.
Root causeOther
ActionOn 03/11/2020, Urgent Product Recall Notices were mailed via second day service. The recalling firm requests that customers immediately stop using, quarantine, and return the affected lots. Further, they requested that customers make personnel within their organization aware of this field action. If the product was further distributed to other facilities, institutions, or manufacturers, customers were asked to ensure this notice was immediately shared with them and note the quantity distributed on the customer response form. Customers with additional questions were encouraged to call 801-208-
Quantity1932
DistributionU.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180723PreludeSYNC DISTAL Radial Compression System
Merit Medical Systems, Inc., Apr 17, 2018