Recall Z-1700-2020 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

Abbott

Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.

DateFeb 28, 2020
StatusTerminated
ReasonInternal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.
Root causeUnder Investigation by firm
ActionOn 02/28/2020, the frim sent all US consignees an "IMPORTANT MEDICAL DEVICE CORRECTION" letter via USPS three to five-day ground service. The customer notification informed customer that they had recently deployed a Patient Care System (PCS) programmer software that contained an anomaly that may result in specific programming changes not being stored in the device. The Recalling Firm explained that when using the affected programmer software to make certain programming changes, the changes may appear to have been successfully programmed but the device could remain at previously programmed se
QuantitySoftware Version 24.6.1 installed into 4733 programmers
DistributionU.S. AK, Al, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. OUS: None - the firm stated that there are no OUS consignees associated to ICDs that were interrogated from January 22, 2020 to F
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.