Recall Z-1804-2020 ยท DQK Computer, Diagnostic, Programmable

Boston Scientific Corporation

iLab Polaris Multi-Modality Guidance System; iLab Ultrasound Imaging System

DateSep 18, 2019
StatusCompleted
ReasonCertain iLab Polaris Systems which were upgraded with a previous 2.10 software version (disk batch 629774) contained a bug in which only the True Color Secondary Capture (SC) format will archive Diastolic Hyperemia-Free Ratio data as intended. If either Multi-Frame UltraSound (US) or SC Image Storage formats are used, the DFR data will not be archived and the user will not be notified of this unexpected behavior.
Root causeOther
ActionOn September 18, 2019, customers that were upgraded with version 54 of iLab Polaris 2.10 (disk batch 629774) software were sent a customer letter via overnight mail delivery. The letter provides instructions and recommendations to the user until a software upgrade (disk batch 636991) was completed. The firm informed that a firm representative will be in contact with customers directly to schedule a software upgrade at their facilities.The firm advises that systems can continue to be used during the time following the recommendations and instructions within the letter. Additionally, in the f
Quantity150 units
DistributionOUS: Japan, Belgium, Denmark, France, Great Britain, and Spain. No U.S. consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191008iLab Polaris Multi-Modality Guidance System
Boston Scientific Corporation, Jul 2, 2019