Recall Z-1753-2020 ยท MSS Folders And Injectors, Intraocular Lens (Iol)

Rayner Intraocular Lenses Limited

RaySert PLUS RSP01. LOT K030. Expiry of 2024-01 - Product Usage: The single use injectors (Model STWOl and Model RSPOl) are intended to be used to compress and insert into the capsular bag only those intraocular lenses that allow the use of these injectors in their approved labelling.

DateMar 17, 2020
StatusOpen, Classified
ReasonRaySert PLUS RSP01 LOT K030 labeled with an expiry of 2024-01 contained an inner tray that was labeled with an expiry of 2015. In addition to having an expiry of 2015 the LOT number on the inner tray is identified as V001.
Root causeLabeling mix-ups
ActionRayner lntraocular notified customers on 03/17/2020 via "Urgent Medical Device recall" letter. The recall letter identified the affected product and lot number. The customers were asked to immediately quarantine and return the product.
Quantity2425 Units.
DistributionUS nationwide distribution in the states of OH, FL, GA, MD, SC, NJ, PR, IN, ID and MN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141091RAYSERT PLUS INJECTOR
Rayner Intraocular Lenses , Ltd., May 28, 2014