Recall Z-1688-2020 ยท JIT Calibrator, Secondary

Abbott Laboratories

"HEMOGLOBIN A1c CALIBRATORS *** REF 4P52-01 *** 307261/R03 *** S4P5X0 *** FOR USE WITH ARCHITECT" - Product Usage: use in the calibration of the Hemoglobin A1c assay.

DateMay 22, 2019
StatusTerminated
ReasonCalibrator level 2 bottle may have been manufactured incorrectly for a portion of the lot. The manufacturer observed no impact to assay performance between the HbA1c concentration range of 5.33% to 6.87%. However, outside of this range, a bias of greater than 3% may be seen with patient results. The device failure that is associated with this recall can result in potentially negatively and positively biased patient test results for HbA1c. The firm is investigating the root cause.
Root causeUnder Investigation by firm
ActionOn May 22, 2019 the firm issued the Product Recall letter along with a customer reply form to all customers informing them that the firm has identified that the calibrator level 2 may have been manufactured incorrectly for a portion of the lots and that although this lot may generate an active calibration curve, a shift to QC and patient results may occur. Between concentration range of 5.33% to 6.87% there is no impact to assay performance, however, outside of this range, a bias of greater than 3% may be seen with patient results. For US customers the Product Recall Letters were sent via Fe
Quantity198 Kits
DistributionWorldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, FKL GA, IA, ID, IL, IN, KS, LA, MN, MO, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, and WI; and countries of Austria, Germany, Italy, South Africa, and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140654Hemoglobin A1c Assay, Hemoglobin A1c Calibrators, Hemoglobin A1c Controls