Recall Z-1688-2020 ยท JIT Calibrator, Secondary
Abbott Laboratories
"HEMOGLOBIN A1c CALIBRATORS *** REF 4P52-01 *** 307261/R03 *** S4P5X0 *** FOR USE WITH ARCHITECT" - Product Usage: use in the calibration of the Hemoglobin A1c assay.
| Date | May 22, 2019 |
| Status | Terminated |
| Reason | Calibrator level 2 bottle may have been manufactured incorrectly for a portion of the lot. The manufacturer observed no impact to assay performance between the HbA1c concentration range of 5.33% to 6.87%. However, outside of this range, a bias of greater than 3% may be seen with patient results. The device failure that is associated with this recall can result in potentially negatively and positively biased patient test results for HbA1c. The firm is investigating the root cause. |
| Root cause | Under Investigation by firm |
| Action | On May 22, 2019 the firm issued the Product Recall letter along with a customer reply form to all customers informing them that the firm has identified that the calibrator level 2 may have been manufactured incorrectly for a portion of the lots and that although this lot may generate an active calibration curve, a shift to QC and patient results may occur. Between concentration range of 5.33% to 6.87% there is no impact to assay performance, however, outside of this range, a bias of greater than 3% may be seen with patient results. For US customers the Product Recall Letters were sent via Fe |
| Quantity | 198 Kits |
| Distribution | Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, FKL GA, IA, ID, IL, IN, KS, LA, MN, MO, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, and WI; and countries of Austria, Germany, Italy, South Africa, and Spain. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K140654 | Hemoglobin A1c Assay, Hemoglobin A1c Calibrators, Hemoglobin A1c Controls Abbott Laboratories, May 22, 2015 |