Recall Z-1995-2020 ยท HAW Neurological Stereotaxic Instrument
Medtronic Navigation, Inc.
Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS (deep brain stimulation) procedure. Neurological stereotaxic Instrument.
| Date | May 18, 2020 (initiated Aug 9, 2019) |
| Status | Terminated |
| Reason | Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the |
| Root cause | Software design |
| Action | Urgent Medical Device Correction Letters were sent to U.S. customers via FedEx Overnight with signature required at delivery beginning on August 9, 2019. European consignees were first notified via registered mail and/or personal visits by Medtronic representatives starting August 27, 2019. The one Canadian consignee was first notified via registered mail starting August 19, 2019. Customers were informed of which devices, when used in combination, can result in entry point and lead placement inaccuracies during Deep Brain Stimulation (DBS) lead implantation procedures. Customers were instruct |
| Quantity | 61 customers purchased the specified combination of equipment which may lead to this error. |
| Distribution | Worldwide distribution including US nationwide, Austria, Canada, Italy, Romania, Spain, and United Kingdom. |