Recall Z-1766-2020 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips North America, LLC

TRx4851A 2.4 GHz IntelliVue Tele TRX, Model Number 862231; Includes Service Numbers 453564052401, 453564052411, 453564052441, 453564052451, 453564166851, 453564166861

DateApr 23, 2020 (initiated Mar 5, 2020)
StatusTerminated
ReasonThe ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.
Root causeSoftware design
ActionUrgent - Medical Device Recall notification letters dated 3/5/20 were sent to customers. " Confirm whether PIC iX Release C.02.xx and C.03.01 is used anywhere in your facility, and if so: o Identify Philips TRx4841A and TRx4851A Telemetry Transceivers in your facility by using the instructions provide in the HOW TO IDENTIFY AFFECTED PRODUCTS section above. o Immediately discontinue use of the Philips TRx4841A and TRx4851A Telemetry Transceiver. Philips support for these devices ended in 2017. " Complete, sign and return the reply form provided on the last page of this letter, as directed on
Quantity21,672
DistributionThe products were distributed US nationwide. Global distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041741M3290A INTELLIVUE INFORMATION CENTER SOFTWARE RELEASE F.0 AND M4840A PHILIPS TELEMETRY SYSTEM II INCLUDING M4841A PATIEN
Philips Medical Systems, Inc., Jul 21, 2004