Recall Z-1886-2020 ยท FPA Set, Administration, Intravascular

CME America, LLC

BodyGuard Microset For Use with the BodyGuard Infusion Pump *** Tubing: 110 in. (280 cm) - Blue Stripe - Product Usage: The system is intended for patients who require maintenance medications, PCA therapy, parenteral nutritional fluids, and general IV fluid therapy in hospital and home care environments.

DateJan 14, 2020
StatusTerminated
ReasonThe filter installed onto the tube set may have the incorrect pore size.
Root causeProcess control
ActionOn 12/20/2019, the firm contacted its distributors via phone call to request that the affected lots not be further distributed pending the Recalling Firm's investigation and assessment for recall. On 01/15/2020, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification to customers via email and FedEx informing them that the filter installed onto the tube set may have the incorrect pore size: -For the lot 90829, the specified size on the label is 0.2 microns. A portion of the tube sets were incorrectly packaged and have a pore size of 1.2 microns rather than 0.2 microns. -For the lot 162
Quantity5,800 infusion sets from lot 90829 and 24,500 from lot 16201
DistributionUS Nationwide distribution including in the states of AK, CA, IL, KS, MA, MO, and SD. No O.U.S.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031749BODYGUARD INFUSION SYSTEM
Caesarea Medical Electronics , Ltd., Aug 21, 2003