Recall Z-1684-2020 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer Medical ApS

ABL800 FLEX, model number/UDI 393-800/0570 In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) In vitro testing of samples of expired air for the parameters pO2 and pCO2 In vitro testing of pleura samples for the pH

DateMar 10, 2020
StatusTerminated
ReasonAnalyzer's barcode reader misinterprets the contents of barcode label used for entering patient ID or accession number. The issue is related to barcode types not using a check digit. This could result in patient mixup or loss of sample resulting in delayed medical treatment
Root causeSoftware design
ActionOn 3/10/2020, a "Urgent: medical Device Recall" letter was sent via email and FedEx to affected consignees. In addition to informing the consignee about the recall issue, the recall notification asked customers to take the following action: Check if your institution is using barcode types without a check digit, including e.g. patient ID or accession number, to be read on the ABL800. 1. If you use barcode types without a check digit, Radiometer recommends to either: - Enable the check digit on the barcode type currently used, or - Change type of barcode to one that includes a check digit, as t
Quantity3,656 units
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, BM, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and countries of Canada, China, Czech Republ
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041874ABL800 FLEX
Radiometer Medical Aps, Oct 8, 2004