ClearanceBrief
Recall Z-1699-2020 ยท GGP Test, Qualitative And Quantitative Factor Deficiency

Diagnostica Stago, Inc.

STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact

DateMar 4, 2020
StatusTerminated
ReasonA risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be falsely increased
Root causeNonconforming Material/Component
ActionStago notified on 3/4/20 end users of the risks to factor VIII tests in writing. Letters issued via email, US mail, and/or UPS overnight mail. Letter states reason for recall, health risk and action to take: To remove this cross-contamination risk, a special pre-wash has been developed for the concerned Stago methodology. It involves special washing of needle 1 before every factor VIII assay in order to eliminate any residual quantities of Emicizumab. This cleaning sequence has been implemented in the stated available software versions. Customers not possessing these software levels will be
Quantity3,225 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K892859THE STAGO VIII DEFICIENT PLASMA KIT
American Bioproducts Co., Jun 23, 1989