Recall Z-1836-2020 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer Medical ApS

ABL800 FLEX model number 393-800 and 393-801, UDI:(01)05700693938004

DateMar 6, 2020
StatusOpen, Classified
ReasonThrough internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay could result in a range of risk from cyanosis and low blood pressure to cardiac arrhythmias and severe impairment of cardiac function.
Root causeComponent design/selection
ActionOn 3/5/2020 a "Urgent: Medical device Recall" letter was hand delivered to effective consignees. In addition to providing information on the recall products, the customer notification as consignees to take the following actions: Solution provided by Radiometer: 1. Your Radiometer representative will exchange the power supply of your ABL800 analyzer at the earliest convenience. Your Actions: 1. Cease using the affected ABL800 analyzer and put it on quarantine. 2. Shut down the affected ABL800 analyzer and disconnect the power cord. 3. Complete the Recall Response Form (last page of this lette
Quantity2 units
DistributionUS: in the state of California
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041874ABL800 FLEX
Radiometer Medical Aps, Oct 8, 2004