Recall Z-1852-2020 ยท NHA Abutment, Implant, Dental, Endosseous

Nobel Biocare Usa Llc

Temporary Abutment Non-Engaging CC RP pre-manufactured prosthetic components directly connected to the endosseous implant, catalog no. 36662 - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function.

DateAug 9, 2019
StatusTerminated
ReasonPackages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct screw.
Root causeProcess control
ActionUrgent Medical Device Recall Letters were sent to U.S. customers via the U.S Postal Service and OUS customers via email beginning on August 9, 2019. This communication notified customers that the firm had identified a packaging issue affecting six product codes and a total of 7 lots of medical devices (titanium abutments) where the packages of abutment lots may contain an incorrect screw. Customers were informed that while the screw will tighten, it cannot be guaranteed that the screw will not loosen. If the screw loosens, customers are instructed to retighten the abutment with the correct
Quantity236
DistributionWorldwide distribution - US Nationwide distribution including in the states of AZ, AR, CA, CO, FL, GA, IL, IN, ID, KS, LA, MI, MD, ME, MA, MO, NY, NC, NJ, NE, NM, OH, OR, OK, PA, TX, TN, VA, WY, WA and WI. Countries of CA, CH, DE, ES, FR, GB, HU, IT, BR, CO, MX, CN, HK, IN, JP and TW.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974150PROCERA PREPARABLE ABUTMENT SYSTEM
Nobel Biocare Uas, Inc., Jan 23, 1998