Recall Z-1904-2020 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Ortho Development Corporation

Tibial Tray Nonporous A Size 4; Model # PN 162-1400-A Material: Ti6Al4V.

DateMay 30, 2019
StatusOpen, Classified
ReasonThe locking mechanism of the tibial tray exhibited a manufacturing defect in which a uniform ridge was cut along the inner wall of the "T" and "Y" junctions, which may cause resistance when the Tibial Insert was being assembled.
Root causeNonconforming Material/Component
ActionThe firm notified their US sales managers via teleconference on 07/19/2019. The sales managers coordinated distribution of the notification letter to the distributors and surgeons. All 12 US surgeons received both in-person verbal and written notices (hand delivered by sales reps) subsequent to the 07/19/2019 teleconference meeting. Recall actions for products sold in Japan were coordinated by the firm's parent company. In the firm's "Urgent Advisory Notice/Medical Device Recall Notice", dated 07/16/2019 includes the following instructions for end users: Contacting affected patients should
Quantity29
DistributionUS: CA, FL, OH, PA, TX, and UT. OUS: Japan.