Recall Z-2032-2020 ยท MCM Implant, Cochlear

Advanced Bionics, LLC

AB HiResolution Bionic Ear System, HiRes Ultra 3D CI, HiFocus SlimJ Electrode - Product Usage: The device is intended to restore a level of auditory sensation to patients with profound sensorineural deafness via electrical stimulation of the auditory nerve.

DateFeb 17, 2020
StatusTerminated
Reasonhearing performance degradation due to body-fluid entering the device.
Root causeDevice Design
ActionOn 02/17/2020, the firm sent an "Urgent Field Action Notification" via email to healthcare providers informing them that the firm has detected an increase in the number of devices being explanted or the potential of being explanted due to performance issues (impedance drop and reports of hearing performance degradation) resulting from fluid ingress at the electrode. The Healthcare Notification instructs Healthcare Providers to: 1. Provide registration information for all devices that have been implanted; 2. Return any unimplanted devices and to emphasize not to implant any of these units. The
Quantity8159 units (CI-1601-04 = 3827 units and CI-1601-05 = 4332 units)
DistributionWorldwide distribution - US Nationwide distribution and the countries of Canada, Brazil, Bangladesh, Switzerland, China, Australia, India, South Korea, Thailand, Ecuador, Mexico, El Salvador, Guatemala, Malaysia, Myanmar, Taiwan, Chile, Costa Rica, Colombia, Panama, Japan, Argentina, Peru, Singapore