Recall Z-2358-2020 ยท PER Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis

Becton, Dickinson and Company, BD Biosciences

BD FACS Sample Prep Assistant (SPA) III - Product Usage: Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II and BD FACSCalibur flow cytometry systems. Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto II flow cytometry systems

DateJul 8, 2019
StatusTerminated
ReasonThe recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.
Root causeComponent design/selection
ActionThe recall was initiated by customer letter dated July 8th, 2019 via UPS/FEDEX and email. The following actions are to be taken: Follow the FACS Sample Prep Assistant II or III System Instructions for Use (IFU) directions for inspecting the probe daily for wear as part of the daily startup procedure. For both SPA II and III IFUs, this may be found in Chapter 5. Instructions for replacing the probe can be located in the Shutdown and Maintenance chapter of the IFU. For SPA II this may be found in Chapter 7. For SPA III this may be found in Chapter 6. The following provides additional detail r
Quantity467 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of AL, CO,CT, GA, IL, MA, MI, MD, NC, NH,NJ, NY, OH, PA, TN, TX, VA, and VT. The countries of Australia, Botswana, Brazil, Canada, Chile, China, Colombia, Costa Rica, Europe, Hong Kong, India, Italy, Jamaica, Japan, Ka
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131301BD FACS SAMPLE PREP ASSISTANT III
Becton, Dickinson & CO, Aug 2, 2013