Recall Z-1672-2020 ยท CDZ Radioimmunoassay, Testosterones And Dihydrotestosterone

Qualigen Inc

The FastPack System Comlete Immunoassay Kit Immunoassay for the Quantitative Measurement of SHBG is a chemiluminescent immunoassay intended for the quantitative determination of Sex Hormone Binding Globulin in human serum and plasma on the FastPack System. The FastPack IP SHBG assay is intended for use as an aid in the diagnosis of androgen disorders. Each Kit Contains 30 FastPack IP SHBG Immun

DateNov 6, 2019
StatusTerminated
ReasonThe chemiluminescent immunoassay has been noted to have higher than expected results which may result in inaccurate quantitative determinations of Sex Hormone Binding Globulin in human serum and plasma.
Root causeNonconforming Material/Component
ActionOn October 31, 2019, a "Urgent: Product Recall" letter was sent to customer via E-Mail. In addition to informing consignees about the recall the customer notification asked consignees to take the following actions: 1. The firm will provide replacement kits at no cost as soon as they become available. 2. For any kits that have been purchased a refund can be requested in place of replacement kits. 3. As part of the product replacement process, please confirm that any remaining SHBG FastPacks/Kits of Lot Number 1908038-1P have been discarded on the form below. 4. The firm will call to confirm
Quantity24 Kits (30 test packs each) 720 tests
DistributionUS Nationwide distribution in the states of AL, AK, CA, HI, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182521FastPack IP Sex Hormone Binding Globulin Immunoassay
Qualigen, Inc., Jan 29, 2019