Recall Z-1508-2020 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Stryker Corporation

Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051971 These products are distributed with the following sheath/probe: 0250070441-PKG, StrykeFlow Electrocautery Probe, Spatula Tip 5MM 0250070442-PKG, StrykeFlow Electrocautery Probe, J Tip, 5MM 0250070443-PKG, StrykeFlow Electrocautery Probe, L Tip, 5MM 0250070444-PKG,

DateSep 12, 2019
StatusTerminated
ReasonPotential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.
Root causeOther
ActionOn September 25, 2019 and November 26, 2019, the firm sent an "Urgent Medical Device Recall Notification" letter to all consignees via USPS mail. In addition, to informing consignees about the recall, they recall letter asked consignees to take the following actions: Actions to be taken by the Customer/User: 1. Inform individuals within your organization who need to be aware of this device removal. 2. Check all stock areas and/or operating room storage to determine if any devices with the affected Strykeprobe Sheath lot numbers from Attachment A are at your facility. - a. If you would like a
Quantity525 devices
DistributionOUS: Netherlands, Spain, Switzerland, Colombia, Canada, Japan, Taiwan,, Puerto Rico, South Africa, and South Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K963765STRYKER STRYKEFLOW ELECTROCAUTERY PROBES
Stryker Endoscopy, Jan 31, 1997