Recall Z-1512-2020 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Boston Scientific Corporation

Boston Scientific, Implantable Cardioverter Defibrillator, DYNAGEN EL ICD DR, Model D152 - Product Usage: Implantable cardioverter defibrillators (ICDs) are intended to provide ventricular anti-tachycardia pacing (ATP) and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.

DateMar 3, 2020 (initiated Feb 12, 2020)
StatusTerminated
ReasonBoston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved by the US Food and Drug Administration.
Root causeUnder Investigation by firm
ActionThe firm initiated the recall by letter on 12/12/2020. The letter explained the error and requested the consignee segregate the units for removal. The firm will provide another model of programmer to be used in the interim.
Quantity1 device
DistributionUS Nationwide distribution in the states of NC, TN, and MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.