ClearanceBrief
Recall Z-1527-2020 ยท OUT Intracranial Aneurysm Flow Diverter

Micro Therapeutics Inc, Dba Ev3 Neurovascular

ev3 Pipeline Flex Embolization Device. For neurological endovascular use.

DateMar 9, 2020 (initiated Feb 14, 2020)
StatusTerminated
ReasonEmbolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex embolization device has already been implanted successfully, there is no increased risk to patients due to this issue.
Root causeProcess change control
ActionOn 02/14/20, Medtronic mailed Urgent Medical Device Recall notices to Healthcare Professionals and Risk Managers. They were advised of the following: 1) Do not use any affected product. Remove and quarantine all unused affected products in your inventory. 2) Return the affected products. Medtronic representatives can assist with facilitating the return of affected product and with identifying suitable replacement product. 3) Complete and return the Customer Confirmation Form. In addition, customers were asked to share the communication within their organization, with other organizations wher
Quantity1050
DistributionWorldwide distribution. US nationwide, Argentina, Australia, Austria, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Italy, Japan, Jordan, Kazakhstan, Korea, Lebanon, Mexico, Netherlands, Nor
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.