Recall Z-1918-2020 ยท FMG Stopcock, I.V. Set
ICU Medical, Inc.
REF: Z2850, 112" (284 cm) 10 Drop Primary set w/2 Clave, Remv 3 Gang 1o2 Manifold w/3 Clave, Baseplate, Rotating Luer w/Filter Cap, 1 Ext, Sterile R, UDI: (01)00887709028241
| Date | Oct 4, 2019 |
| Status | Terminated |
| Reason | There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences. |
| Root cause | Mixed-up of materials/components |
| Action | On October 4, 2019, a "Urgent: Medical Device Recall" Letter was sent to all affected customers via US Mail. In addition to providing customers information on the recalled products, the letter asked customers to do the following: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, |
| Quantity | 25 units |
| Distribution | US:AK,AL,AR,AZ,CA,CO,CT,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,SD,TN,TX,UT,VA,WA,WI,WY OUS: Canada *** "REF" numbers AH7117 and B30211 were not marketed or distributed within the US*** |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K964435 | PRIMARY IV SET Icu Medical, Inc., Apr 2, 1997 |