Recall Z-1497-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis zee Biplane, Model No. 10094141; Artis zee Ceiling, Model No. 10094137; Artis zee Floor, Model No. 10094135; Artis zeego, Model No. 10280959 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

DateMar 13, 2020 (initiated Feb 14, 2020)
StatusTerminated
ReasonA software issue could potentially cause the stand and table movements to be blocked.
Root causeSoftware design
ActionOn February 14, 2020, the firm distributed Urgent Medical Device Correction letters to customers. Customers were informed of the software error which affects the listed systems. In case of software failure, planned procedures may have to be terminated and performed on an alternative x-ray system. Customers were instructed to do the following: Issue # 1: Potential issue with the Message "Stand battery charging: DO NOT POWER OFF" on Artis zeego systems Customer Action: Please check the system messages before power off/shutdown of the system. If possible, do not switch off the system while the
Quantity1452 worldwide (638 US)
DistributionUS Nationwide distribuiton.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018