Recall Z-1682-2020 ยท DQK Computer, Diagnostic, Programmable

CHANGE HEALTHCARE CANADA COMPANY

McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management

DateFeb 10, 2020
StatusOpen, Classified
ReasonInaccurate estimation of the heart rate may occur in instances where the user is estimating heart rate counting grid squares using the Sequential ECG format
Root causeDevice Design
ActionChange Healthcare Canada Company notified customers on about 02/10/2020 via "Urgent Field Safety Notice" letter. Customers were informed of an issue whereby users who are estimating heart rate by counting grid squares, may assume that the Sequential ECG format is presented on a scale of 25 mm/sec and thereby may inaccurately estimate the heart rate value. Instructions included that the Simultaneous ECG format can be used rather than the Sequential ECG format, until the software is updated and the Sequential ECG format option can be used while maintaining awareness to the scale bar on the s
Quantity51 units
DistributionDistribution US nationwide, Canada and the United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113515HORIZON CARDIOLOGY ECG MANAGEMENT
Mckesson Israel , Ltd., Mar 16, 2012