ClearanceBrief
Recall Z-1846-2020 ยท DXE Catheter, Embolectomy

Applied Medical Resources Corp

Python Embolectomy Catheters, Sterile EO, CE, Ref numbers/UDI: A4E01/(01)00607915113407; A4E02/(01)00607915113414; A4E03/(01)00607915113421; A4E04/(01)00607915113438; A4E05/(01)00607915113445; A4E06/(01)00607915113452; A4E08/(01)00607915113476; A4E09/(01)00607915113483

DateMay 7, 2020 (initiated Oct 24, 2019)
StatusTerminated
ReasonThe firm has received reports of tip separation during usage of the catheters, which may expose patients to unintended dislodgement of the tip from the body of the catheter.
Root causeComponent change control
ActionOn November 08, 2019, a "Urgent: Medical Device Recall" letter was sent via UPS to all affected customers. In addition to providing information on the recalled products, the letter asked customers to do the following: 1. Check your inventory for the recalled product. 2. Complete the Recall Notification Confirmation Form to acknowledge the recall and indicate if your facility is returning or has already used product from the lot listed above. 3. If no product is being returned, please indicate on the Recall Notification Confirmation Form. 4. If you are a distributor, please notify any fac
Quantity28,799 catheters in total
DistributionWorldwide distribution - United States: Nationwide OUS: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Poland, Saudi Arabia, South Africa, Spain, Taiwan, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K970762DUAL LUMEN EMBOLECTOMY CATHETER
Applied Medical Resources, Mar 28, 1997