Recall Z-1540-2020 ยท CIG Diazo Colorimetry, Bilirubin

Siemens Healthcare Diagnostics, Inc.

Dimension Total Bilirubin Flex reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10444957 Product Usage: Dimension TBI method for the clinical chemistry syst6em is an in vitro diagnostic test intended to quantitatively measure total bilirubin in human serum and plasma. Dimension Measurements of total bilirubin are used in the diagnosis and treatment of liver

DateFeb 13, 2020
StatusTerminated
ReasonBilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
Root causeUnder Investigation by firm
ActionSiemens Healthineers issued Urgent Medical Device Correction (UMDC # ACHC-20-02.A.US, ACHC-20-02.A.US.AC, ACHC-20-02.A.US.DM, ACHC-20-02.A.US.DV) and Urgent Field Safety Notice (UFSN # ACHC-20-02.A.OUS, ACHC-20-02.A.OUS.AC, ACHC-20-02.A.OUS.DM, ACHC-20-02.A.OUS.DV) on February 13, 2020 to all Atellica CH, ADVIA CH, Dimension, and Dimension Vista customers to inform them of the issue, the potential risk to health, and actions to be taken by the customer. Customers were instructed to:Review the information on eltrombopag interference (ADVIA CH and Atellica CH bilirubins) or use of total bilir
Quantity1639 US and 747 OUS
DistributionNationwide Foreign: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, Un
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060628DIMENSION TBI FLEX AND DBI FLEX REAGENT CARTRIDGES AND TBI/DBI CALIBRATOR
Dade Behring, Inc., Apr 6, 2006