Recall Z-1474-2020 ยท FRN Pump, Infusion

CME America, LLC

CME America BodyGuard Infusion Pump System, All models of BG 323, BG 121, BG 545, BG CV545, BG 575, BG CV575, and CMExpress

DateMar 18, 2020 (initiated Jan 7, 2020)
StatusTerminated
ReasonInfusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing at a very low rate (0.1 mL/h), or 2) slower than expected drug delivery when infusing at high flow rates (greater-than 500 mL/h)
Root causeDevice Design
ActionOn 01/07/20, Urgent Medical Device Correction notices were emailed to distributors and mailed to direct customers. For direct customers, Recommendation on further product use included: The recalling firm recommends that customers continue to use the infusion pump system as per intended use. 1. As per standard clinical practice, clinicians administering infusions should assess the fluid container for volumes infused, volumes remaining in the container at the end of the infusion and ensure the total volume of prescribed medication is delivered. 2. Verify your maintenance records to confirm you
Quantity30,596 pumps
DistributionU.S.: HI, OR, AR, MA, PA, OH, WY, CO, NE, NC, WA, TX, UT, MD, CA, KY, FL, NH, TN, NY, SD, MO, IL, WI, MN, RI, VA, DC, SC, NJ, ME, KS, CT, MI, LA, AK, NM, IA, MT, ID, AZ, AL, VT, IN, MS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061325BODYGAURD PAIN MANAGER INFUSION SYSTEM
Caesarea Medical Electronics , Ltd., Aug 24, 2006