Recall Z-1502-2020 ยท IZF System, X-Ray, Tomographic

Agfa N.V.

AGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Usage: TomoSynthesis is used to synthesize tomographic slices from a single tomographic sweep. The DR 800 is not intended for mammography applications

DateFeb 10, 2020
StatusTerminated
ReasonTomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.
Root causeUnder Investigation by firm
ActionAFGA issued via email On February 10, 2020, an URGENT FIELD SAFETY NOTICE letter. Letter states reason for recall, health risk, and action to take: With this letter, Agfa is informing you that the DR 800 system still can be used including for Tomosynthesis for mAs values which are larger than 0,2 mAs. In case Tomosynthesis image acquisition sequence does not stop automatically after the expected number of exposures, release the exposure button as this will stop the sequence and notify your Agfa Service contact at once. Contact your local Agfa Imaging organization 1-877-777-2432 and please re
Quantity2 units
DistributionUS Nationwide distribution in the states of FL, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183275DR 800 with Tomosynthesis
Agfa N.V., Feb 1, 2019