Recall Z-1599-2020 ยท DIC Alcohol Dehydrogenase, Specific Reagent For Ethanol Enzyme Method

Siemens Healthcare Diagnostics, Inc.

Atellica CH Ethyl Alcohol, Test Code ETOH, Siemens Material Number (SMN) 11097501, UDI 00630414596112. The product is an ethyl alcohol assay, IVD use.

DateFeb 27, 2020 (initiated Feb 14, 2020)
StatusTerminated
ReasonThe assay may exhibit a positive bias in QC and patient results within the 10-day pack calibration interval.
Root causeUnder Investigation by firm
ActionOn February 11, 2020, the firm issued an Urgent Medical Device Correction to all US customers. Customers who experience the issue were instructed to perform a pack calibration every 2 days or more frequently based on the laboratory's Quality Control performance. The firm is working to address the issue in a future software version. Customers are requested to complete and return the effectiveness check questionnaire. If you have any questions, contact your local Siemens Healthineers technical support representative.
Quantity1124
DistributionDomestic distribution nationwide. Foreign distribution to Australia, Austria, Bahamas, Belgium, Canada, Curacao, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kenya, Latvia, Mexico, Netherlands, Norway, Poland, Portugal, Slovakia, South Korea, Spain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071811ETOH FLEX REAGENT CARTRIDGE (DF22)
Dade Behring, Inc., Aug 24, 2007