Recall Z-1554-2020 ยท KMI Monitor, Bed Patient

Hill-Rom Inc.

NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Software is compatible with the following beds: 1. VersaCare Revision A-J with upgrade kit Revision K with UCB 1.18 2. Centrella (all versions) 3. Progressa (firmware version 1.15.0.0)

DateFeb 14, 2020
StatusTerminated
ReasonThere is a software issue which may result in failure to monitor or control patient's bed exit mechanism.
Root causeSoftware Manufacturing/Software Deployment
ActionThe firm issued a Urgent Medical Device notification on 02/14/2020 by letter. The letter identified the reason and requested the following actions be taken: 1. Hillrom is recommending customers disable the NPS software module, including monitoring for Bed Exit Not Armed, to avoid the potential for staff being unaware of the bed exit state. 2. With the NPS software disabled, caregivers will need to return to manually arming the bed exit to alert for patients exiting the bed. Hillrom is developing a software update for NaviCare Patient Safety and anticipates availability in April 2020. Next S
Quantity35 Software Systems
DistributionColorado, Georgia, Indiana, Louisiana, North Carolina, Nevada, Oregon, South Carolina, Tennessee, Virginia, Washington.