Recall Z-1823-2020 · OVE Intervertebral Fusion Device With Integrated Fixation, Cervical

U&I CORP.

Velofix SA cervical cage, REF: TSA161207, TSA161208, TSA161209, TSA161210, TSA161407, TSA161408, TSA161409, TSA161410 - Product Usage:The device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disk disease at one level from C2-C3 disc to the C7-T1 disc. Each product is composed of four components: PEEK cage, titanium plate, and two titaniu

DateJan 30, 2019
StatusTerminated
ReasonThe vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, lossening, instability, or less bone-screw engagement than intended by physicians.
Root causeEmployee error
ActionOn 01/30/2019, "Notification for Sales Suspension and Product recall" and explanation of nonconformity notices were emailed to the domestic customer. The notification requests customers identify, isolate and return affected products immediately. It requests acknowledgement, and quantity of devices to be returned. The contact number for further questions is +82 31 860-6830 or kkjane@youic.com
Quantity42
DistributionUS Nationwide distribution in the state of CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172424Velofix™ SA Cervical Cage
U&I Corporation, Nov 22, 2017