Recall Z-1579-2020 ยท LJT Port & Catheter, Implanted, Subcutaneous, Intravascular

Bard Peripheral Vascular Inc

PowerFlow Implatable Apheresis IV Port with attachable 9.6 F ChronoFl x Open-Ended Single-Lumen Venous Catheter, Ref A710962, UDI:(01)00801741129438 Product Usage: Bard implantable ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for t

DateOct 1, 2019
StatusTerminated
ReasonFirm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.
Root causeComponent change control
ActionOn October 1, 2019, Bard sent a "Urgent Medical Device REcall Notification" letter to all customers via FedEx. The recall notification informed customer about the reason for the recall. In addition, to informing customers about the recall, the letter asked that customers take the following actions: 1. Our records show that your facility has purchased product codes affected by this voluntary recall. Do not use or further distribute any affected product. 2. Please check all inventory locations within your institution for affected product code / lot number combinations listed in the recall not
Quantity296 Catheters
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS,KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT,VA, VT, WA, WI, WV, WY and countries of Brazil, Canada, EMEA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163001PowerFlow Apheresis I.V. Port
C.R. Bard, Inc., Apr 17, 2017