Recall Z-1513-2020 ยท CHP Radioimmunoassay, Estradiol

Siemens Healthcare Diagnostics, Inc.

IMMULITE 1000 Systems E2 Estradiol Catalog #: US: LKE21(D) and OUS: LKE21 - Product Usage: For in vitro diagnostic use with the IMMULITE systems Analyzers for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders.

DateFeb 24, 2020 (initiated Feb 5, 2020)
StatusTerminated
ReasonDue to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed initiation of a potentially beneficial drug and/or administration of an unnecessary drug in the treatment for hormone receptor positive advanced or metastatic breast cancer.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Correction (UMDC) IMC20-01.A.US was sent to US customers and an Urgent Field Safety Notice (UFSN) IMC20-01.A.OUS was sent to OUS customers in February 2020. The UMDC and UFSN advise customers of the potential for high discordant Estradiol results in some samples on the IMMULITE systems. Siemens Healthineers has determined kit lots 501 and above released in July 2018 for the IMMULITE systems are potentially affected. The UMDC instructs US customers to discontinue use of and discard Estradiol kits currently in their inventory. Customers will also be asked to do a loo
Quantity2267 kits
DistributionWorldwide distribution - US Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, KY, LA, MA, MD, MI, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI. SC, TN, TX, UT, VA, WA, WI, WY, and countries of AE, AF, AL, AO, AR, AT, AU, AZ, BD, BE, BO, BG, BH, BR, CA, CH, CO, CL, CN, CO,