Recall Z-1524-2020 ยท OSR Pacemaker/Icd/Crt Non-Implanted Components

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Medtronic CareLink 2090 Programmer. Used to interrogate and program Medtronic and Vitatron implantable devices, such as pacemaker/ICD/CRT.

DateJan 30, 2020
StatusOpen, Classified
ReasonMedtronic Conexus Telemetry has been determined to contain two primary cyber vulnerabilities: improper access control and the cleartext transmission of sensitive information.
Root causeSoftware design
ActionMedtronic initially notified customers on about 03/21/2019 via a "Security Bulletin" regarding vulnerabilities to the proprietary Medtronic Conexus radio frequency wireless telemetry protocols associated with some Medtronic ICDs (implantable cardioverter defibrillators) and CRT-Ds (cardiac resynchronization therapy defibrillators). On about 01/30/2020, Medtronic notified customers with affected programmers of the field corrective action notification via 2-day UPS delivery of the consignee letter. On 21 March 2019, Medtronic posted a Security Bulletin regarding the vulnerabilities. Custome
Quantity23,180 devices
DistributionWorldwide distribution. US Nationwide including Puerto Rico, Argentina, Brazil, Colombia, Trinidad and Tobago, Venezuela, Dominican Republic, Bahamas, Peru, Bermuda, Barbados, Ecuador, Cayman Islands, Curacao, Guatemala, Honduras, Panama, El Salvador, Costa Rica, Bolivia, Chile, Uruguay, Paraguay,
FDA recordOpen on fda.gov