Recall Z-1434-2020 ยท KNW Instrument, Biopsy
Bard Peripheral Vascular Inc
Bard Biopsy EnCor Probe, MRI, 10G Blunt Tip, REF number ECPMR0110GBT, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile.
| Date | Jan 3, 2020 |
| Status | Terminated |
| Reason | Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples. |
| Root cause | Component change control |
| Action | The recalling firm issued letters dated 1/3/2020 to each of the U.S. consignees via FedEx with proof of delivery notification. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-321-4254 Option #5. |
| Quantity | 1,085 units |
| Distribution | Distribution was made to AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, OR, PA, SC, TX, VA, and WV. There was government distribution and no military distribution. Foreign distribution was made to Canada, Australia, Austria, Belgium, China, Czech Re |
| FDA record | Open on fda.gov |