Recall Z-1433-2020 ยท KNW Instrument, Biopsy

Bard Peripheral Vascular Inc

Bard Biopsy EnCor Probe, MRI, 10G Trocar Tip, REF number ECPMR0110G, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile.

DateJan 3, 2020
StatusTerminated
ReasonLot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples.
Root causeComponent change control
ActionThe recalling firm issued letters dated 1/3/2020 to each of the U.S. consignees via FedEx with proof of delivery notification. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-321-4254 Option #5.
Quantity2,050 units
DistributionDistribution was made to AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, OR, PA, SC, TX, VA, and WV. There was government distribution and no military distribution. Foreign distribution was made to Canada, Australia, Austria, Belgium, China, Czech Re
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040842SENORX BIOPSY DEVICE II
Senorx, Inc., Apr 30, 2004
K051158ENCOR 7 GAUGE
Senorx, Inc., May 16, 2005