Recall Z-1741-2020 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352.

DateMay 17, 2019
StatusTerminated
ReasonPotential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.
Root causeSoftware design
ActionThe recalling firm issued letters dated 5/8/2019 on 5/17/2019 informing the customers of the issue and that a spilled sample could lead to exposure of biohazardous material and that a software update would be made by their field service engineers at the next service call.
Quantity291 units
DistributionDistribution was made to AL, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, MI, NC, ND, NJ, NY, OH, OR, PA, SD, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Argentina, Australia, Canada, China, Czech Republic, Egypt, France, Germany, In