Recall Z-1444-2020 ยท NIG System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer

Roche Diagnostics Operations, Inc.

Elecsys CA 19-9, Cat. No. 07027028190 - Product Usage: the product is an immunoassay for monitoring and management of pancreatic cancer

DateMar 5, 2020 (initiated Feb 7, 2020)
StatusTerminated
ReasonThe firm has received an increased number of complaints concerning non-reproducible elevated results ("high flyers") for one lot of Elecsys CA 19-9 reagent, which may affect clinical interpretation.
Root causeNonconforming Material/Component
ActionOn February 7, 2020, the firm issued an Urgent Medical Device Correction letter to affected customers. Customers were informed of the product issue and given two options for CA 19-9 testing until another reagent lot is available for the cobas e 801 module: 1) Use a different Roche platform for testing, or 2) Implement a workaround, described in detail in the customer letter. The letter also provided instructions for receiving credit depending on which option was elected. Roche will notify customers when a new lot of CA 19-9 reagent for the cobas e 801 is available for ordering. ***Updated 10
Quantity684
DistributionUS Nationwide distribution in the states of AL, AZ, CA, CO, FL, HI, IL, IN, KS, KY, MA, ME, MI, NC, NH, NJ, NY, TN, TX, WA and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050231ELECSYS CA 19-9 IMMUNOASSAY, ELECSYS CA 19-9 CALSET
Roche Diagnostics Corp., Jul 6, 2005