Recall Z-1843-2020 ยท ITX Transducer, Ultrasonic, Diagnostic
Philips Ultrasound Inc
Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5, EPIQ CVxi Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies
| Date | Aug 30, 2019 |
| Status | Terminated |
| Reason | The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury |
| Root cause | Process control |
| Action | If you have an EPIQ, Affiniti or iE33 system, verify that all your S7-3t and S8-3t TEE transducers are programmed properly by connecting them to the system, checking the transducer temperature (TEE T) on the display, and following the instructions provided on this letter. If you determine that your transducer is not properly programmed, discontinue using the transducer immediately and contact your local Philips representative to schedule an appointment with a field service engineer. If you have an HD11 or CX50 system discontinue using the S7-3t and S8-3t transducer immediately and contact yo |
| Distribution | Worldwide distribution - US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K172607 | Philips EPIQ 5 Diagnostic Ultrasound System, EPIQ 7 Diagnostic Ultrasound System, Affiniti 30 Diagnostic Ultrasound Syst Philips Health Care, Oct 4, 2017 |