Recall Z-1841-2020 ยท ITX Transducer, Ultrasonic, Diagnostic

Philips Ultrasound Inc

Transducer S7-3t UDI (01)00884838061439 REF 989605406772 Can be used with: CX50, EPIQ 7, Affiniti 50 and Affiniti 70 Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies

DateAug 30, 2019
StatusTerminated
ReasonThe firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
Root causeProcess control
ActionIf you have an EPIQ, Affiniti or iE33 system, verify that all your S7-3t and S8-3t TEE transducers are programmed properly by connecting them to the system, checking the transducer temperature (TEE T) on the display, and following the instructions provided on this letter. If you determine that your transducer is not properly programmed, discontinue using the transducer immediately and contact your local Philips representative to schedule an appointment with a field service engineer. If you have an HD11 or CX50 system discontinue using the S7-3t and S8-3t transducer immediately and contact yo
DistributionWorldwide distribution - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160807EPIQ 5 Diagnostic Ultrasound System, EPIQ 7Diagnostic Ultrasound System, Affiniti 50 Diagnostic Ultrasound System, Affin
Philips Ultrasound, Inc., Apr 6, 2016
K162329CX50 Diagnostic Ultrasound System, Sparq Diagnostic Ultrasound System
Philips Health Care, Sep 14, 2016