Recall Z-1460-2020 ยท MWJ Electrocardiograph, Ambulatory (Without Analysis)

Braemar Manufacturing, LLC

Braemar Manufacturing LLC DL950 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only

DateJan 22, 2020
StatusTerminated
ReasonAn issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Err
Root causeSoftware design
ActionConsignees notified via email with recall letter dated 01/22/2020. The letter identified affected product, stated reason for recall, and provided instructions on how to reset the error by following provided instructions, or to send the equipment to Braemar for repair.
Quantity10582
DistributionNationwide. No governmental consignees or Canadian consignees. International distribution to Germany, Spain, Switzerland, France, Finland, Vietnam, Israel, France, Hong Kong, Czech Republic, Finland, Belgium, Sweden, Mexico, Lebanon, United Arab Emirates, Bolivia, Italy, and Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071733DL900 SERIES HOLTER RECORDER
Braemar, Inc., Jul 24, 2007