Recall Z-2240-2020 ยท MSW System, Test, Thyroglobulin
Beckman Coulter Inc.
Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin, REF 33860 (UDI: 15099590227173) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.
| Date | Jun 2, 2020 (initiated Apr 30, 2019) |
| Status | Terminated |
| Reason | The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays (Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin. |
| Root cause | Other |
| Action | On 04/30/2019, an Urgent Medical Device Letter was sent via mail to affected consignees. Customers were instructed to access the updated Instructions for Use (IFU) Limitations and Interferences sections for the six affected Access immunoassays on the firm's website. Additionally they ask customers to do the following: 1) Share the information with laboratory staff and retain the notification as part of their laboratory Quality System documentation. If any of the affected product(s) have been forwarded to another laboratory, they are to provide them with a copy of the Urgent Medical Device Rec |
| Quantity | 95,453 units |
| Distribution | US Distribution: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusett, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, |
| FDA record | Open on fda.gov |