ClearanceBrief
Recall Z-1443-2020 ยท BSY Catheters, Suction, Tracheobronchial

RESPIRATORY THERAPEUTICS GROUP LLC

RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 - Product Usage: To be used to aspirate liquids or semi solids from a patient s upper airway.

DateAug 19, 2019
StatusOpen, Classified
ReasonIt has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea.
Root causeComponent change control
ActionOn 8/19/19 a recall notice was sent to all affected customers. The letter outlined the recall issue and requested that customers complete the enclosed form for the return of the products (new and used) and the issuance of replacement product or credit.
Quantity723 cases (7230 units)
DistributionUS Nationwide distribution in the states of CA, NJ, NY. OUS: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall