Recall Z-1351-2020 ยท NVZ Pulse Generator, Permanent, Implantable
St Jude Medical, Cardiac Rhythm Management Division
Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C (05414734507585), CD1277-36 (05414734506045), CD1377-36C (05414734507622), CD2277-36Q (05414734505932), CD2377-36QC (05414734507523), with Merlin Patient Care System with software model 3330
| Date | Jan 22, 2020 |
| Status | Terminated |
| Reason | Past updates to programmers and transmitters may lead some implantable cardioverter defibrillators (ICD) to lose wireless radiofrequency (RF) communication; which means patients can no longer be interrogated with wireless RF telemetry or monitored remotely. |
| Root cause | Other |
| Action | Starting on 01/22/20, Important Medical Device Correction notices were hand delivered by the recalling firm's sales representatives. Physicians were informed that the recalling firm developed a software patch for programmers which restores wireless RF communication capability in affected devices. Device explant is not required for the update. The recalling firm will assist in updating programmer software and restoring wireless RF communication for affected devices. Correction notices recommend working with firm representatives to help correct affected devices during the patient's next regu |
| Quantity | 256 |
| Distribution | U.S.: WV, MS, TX, DE, OH, AZ, MI, NC, KY, WA, NY, CA, IN, MO, MD, NJ, FL, PA, OK. O.U.S.: United Kingdom, Turkey, Sweden, Spain, Portugal, Poland, Philippines, Pakistan, Norway, New Zealand, Netherlands, Japan, Italy, Hong Kong, Greece, Germany, France, Finland, Denmark, Canada, Belgium, Austria, A |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.