Recall Z-1354-2020 ยท NHN Powered Light Based Laser Non-Thermal Instrument With Non-Heating Effect For Adjunctive Use In Pain Therapy

Theralase Inc.

TLC-2000 Therapeutic Medical Laser System

DateFeb 13, 2020 (initiated Sep 26, 2019)
StatusTerminated
Reason905 nm laser diodes may be below rated power output.
Root causeDevice Design
ActionEmail was sent out to healthcare practitioners regarding the recall on September 26, 2019. .Healthcare practitioners are instructed to courier their TLC-2000 to Theralase for analysis. Their 905 nm laser diodes will be analyzed and recalibrated, as required, to meet rated power output.
Quantity92 units
DistributionUSA; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151816TLC-2000 Therapeutic Medical Laser System
Theralase, Inc., Nov 19, 2015