Recall Z-1423-2020 ยท CDT Lipase Hydrolysis/Glycerol Kinase Enzyme, Triglycerides

Beckman Coulter Inc

Triglycerides Reagent (TG 2 x 300), REF: 445850, For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233228

DateAug 15, 2019
StatusTerminated
ReasonIt has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.
Root causeDevice Design
ActionOn August 15, 2019, the firm mailed or emailed a "Urgent Medical device Recall" to affected consignees In addition to informing the consignees about the recall, they ask consignees to do the following: 1. Resolution: The following statement will be added to the Interfering Substances section of the Enzymatic Creatinine, Triglycerides GPO Blanked, Uric Acid, Direct Bilirubin, and Total Bilirubin Chemistry Information Sheets (CIS): N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may generate erroneously low results in samples for patients that have taken to
DistributionAlabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.