Recall Z-1580-2020 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Gmbh & Co. KG

LN 3R70-01; software version 2.6.2 and earlier. The software is used with the Alinity i (LN 03R65-01) processing module and the Alinity c (LN 03R67-01) processing module.

DateJan 24, 2020
StatusTerminated
ReasonAbbott has identified potential performance issues for the Alinity ci -series Software version 2.6.2 and earlier.
Root causeSoftware design
ActionThe firm initiated the recall letter on January 24, 2020. The "Product Correction, Urgent - Immediate Action Required" notice identified the correction of the Alinity ci-series Software version 2.6.2 and earlier due to potential performance issues. the notice informed that Abbott is releasing Alinity ci-series Software version 3.1.0 to correct the issues. The notice provided the necessary actions required to be performed until the mandatory upgrade to SW 3.1.0 is complete. The notice referenced that details are included in table1 below. the notice detail that an Abbott representative will
Quantity2676 units
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, ID, IL, LA, MA, MD, ME, MI, MN, MC, MS, NC, NJ, NY, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and countries of Albania, Argentina, Australia, Austria, Bahrain, Bangledesh, Belgium, Bermuda, Bosnia, Bots
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.