Recall Z-1427-2020 ยท DSQ Ventricular (Assist) Bypass

Heartware, Inc.

HeartWare HVAD System Battery Charger The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.

DateMar 9, 2020 (initiated Jan 23, 2020)
StatusOpen, Classified
ReasonPotential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
Root causeDevice Design
ActionHeartware notified customers on about 01/23/2020 via "Urgent Field Safety Notice" letter. Customers were notified of the potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port. When this occurs, communication circuits between the battery and the Controller may be damaged. If circuit damage occurs, it will be necessary to exchange the Controller to restore communication between the batteries and Controller. Instructions included to advise patients to carefully follow the guidance provided in the patient manual related to conne
Quantity5,489 devices total
DistributionNo affected devices distributed within the US. International distribution to Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, EGYPT, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Korea, Korea, Republic o